Effects of Confidence-Based Rejection on Usability and Error in Pattern Recognition-Based Myoelectric Control

dc.contributor.authorRobertson, Jason W.
dc.contributor.authorEnglehart, Kevin B.
dc.contributor.authorScheme, Erik J.
dc.date.accessioned2023-06-28T13:56:53Z
dc.date.available2023-06-28T13:56:53Z
dc.date.issued2019
dc.description.abstractRejection of movements based on the confidence in the classification decision has previously been demonstrated to improve the usability of pattern recognition based myoelectric control. To this point, however, the optimal rejection threshold has been determined heuristically, and it is not known how different thresholds affect the tradeoff between error mitigation and false rejections in real-time closed-loop control. To answer this question, 24 able-bodied subjects completed a real-time Fitts' law-style virtual cursor control task using a support vector machine classifier. It was found that rejection improved information throughput at all thresholds, with the best performance coming at thresholds between 0.60 and 0.75. Two fundamental types of error were defined and identified: operator error (identifiable, repeatable behaviors, directly attributable to the user), and systemic error (other errors attributable to misclassification or noise). The incidence of both operator and systemic errors were found to decrease as rejection threshold increased. Moreover, while the incidence of all error types correlated strongly with path efficiency, only systemic errors correlated strongly with throughput and trial completion rate. Interestingly, more experienced users were found to commit as many errors as novice users, despite performing better in the Fitts' task, suggesting that there is more to usability than error prevention alone. Nevertheless, these results demonstrate the usability gains possible with rejection across a range of thresholds for both novice and experienced users alike.
dc.description.copyright© 2019 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
dc.identifier.doi10.1109/jbhi.2018.2878907
dc.identifier.issn2168-2194
dc.identifier.issn2168-2208
dc.identifier.urihttps://unbscholar.lib.unb.ca/handle/1882/37236
dc.language.isoen
dc.publisherInstitute of Electrical and Electronics Engineers
dc.relationNSERC
dc.relationDiscovery
dc.relation.hasversionhttp://dx.doi.org/10.1109/JBHI.2018.2878907
dc.relation.ispartofIEEE Journal of Biomedical and Health Informatics
dc.rightshttp://purl.org/coar/access_right/c_abf2
dc.subject.disciplineElectrical and Computer Engineering
dc.titleEffects of Confidence-Based Rejection on Usability and Error in Pattern Recognition-Based Myoelectric Control
dc.typejournal-article
oaire.citation.endPage2008
oaire.citation.issue5
oaire.citation.startPage2002
oaire.citation.titleIEEE Journal of Biomedical and Health Informatics
oaire.citation.volume23
oaire.license.conditionother
oaire.versionhttp://purl.org/coar/version/c_ab4af688f83e57aa

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Robertson et al. JBHI 2019, Final Submission.pdf
Size:
305.71 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.13 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections