On the usability of intramuscular EMG for prosthetic control: A Fitts’ Law approach
Loading...
Date
2014-10
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier
Abstract
Previous studies on intramuscular EMG based control used offline data analysis. The current study investigates the usability of intramuscular EMG in two degree-of-freedom using a Fitts’ Law approach by combining classification and proportional control to perform a task, with real time feedback of user performance. Nine able-bodied subjects participated in the study. Intramuscular and surface EMG signals were recorded concurrently from the right forearm. Five performance metrics (Throughput, Path efficiency, Average Speed, Overshoot and Completion Rate) were used for quantification of usability. Intramuscular EMG based control performed significantly better than surface EMG for Path Efficiency (80.5 ± 2.4% vs. 71.5 ± 3.8%, P = 0.004) and Overshoot (22.0 ± 3.0% vs. 45.1 ± 6.6%, P = 0.01). No difference was found between Throughput and Completion Rate. However the Average Speed was significantly higher for surface (51.8 ± 5.5%) than for intramuscular EMG (35.7 ± 2.7%). The results obtained in this study imply that intramuscular EMG has great potential as control source for advanced myoelectric prosthetic devices.